首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Absolute quantum efficiency calibration of analog detectors with twin photon current
Authors:Dongyang Gao  Maopeng Xia  Jianjun Li  Youbo Hu  Xiaobing Zheng
Institution:1.Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics,Chinese Academy of Sciences,Hefei,China;2.University of Science and Technology of China,Hefei,China
Abstract:Correlated photon is exploited as a tool for absolute quantum efficiency calibration of photon detectors working in the photon-counting regime. We extend the method from photon-counting regime to the absolute calibration of analog regime photon detectors. A revised calibration module is proposed based on balance detection and fluctuation restrictions conditions applied to analog photon detectors. An absolute calibration experimental system is established. This method shows a relative standard uncertainty of 2.1% and a relative deviation of 1% compared with those of the typical photomultiplier quantum efficiency.
Keywords:
本文献已被 CNKI SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏