首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Control system for low pressure gradient mixer in high performance liquid chromatograph
Authors:Ying Zhao  Jun-yan Wang  Qiu-shi Ren  Xin-yu Chai
Institution:ZHAO Ying,WANG Jun-yan,REN Qiu-shi,CHAI Xin-yu(Department of Biomedical Engineering,Shanghai Jiaotong University,Shanghai 200240,China)
Abstract:A control system used in high performance liquid chromatograph(HPLC) was described.The control system adopting low pressure gradient elution was tested with different initial and end volume fractions,and four types of gradient elution curves.The experimental results verified the theoretical analyses of the applied method.This self-designed control system can achieve approving accuracy,repeatability and low cost,which has a bright outlook for domestic applications.
Keywords:low pressure  gradient  high performance liquid chromatograph(HPLC)  mixer  